Site Master™ S331E, S332E, S361E, S362E Specifications
Power Meter (Option 0029) (S332E, S362E)
Frequency Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #, Full Band
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average Acquisition Fast/Med/Slow, # of Running Averages
Limits Limit On/Off, Limit Upper/Lower
Frequency Range 10 MHz to 4 GHz (S332E), 10 MHz to 6 GHz (S362E)
Span 1 kHz to 100 MHz
Display Range –140 dBm to +30 dBm, ≤ 40 dB span
Measurement Range –120 dBm to +26 dBm
Offset Range 0 to +100 dB
VSWR 2:1 typical
Maximum Continuous Input +35 dBm without attenuator
Accuracy Same as Spectrum Analyzer
Application Options Impedance (50 Ω, 75 Ω, Other)
High Accuracy Power Meter (Option 0019) (Requires external USB Power Sensor(s))
Amplitude Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average # of Running Averages, Max Hold
Zero/Cal Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits Limit On/Off, Limit Upper/Lower
Power Sensor Model PSN50 MA24104A MA24106A MA24108/18/26A
Description High Accuracy RF Power
Sensor
Inline High Power Sensor High Accuracy RF Power
Sensor
Microwave USB Power
Sensor
Frequency Range 50 MHz to 6 GHz 600 MHz to 4 GHz 50 MHz to 6 GHz 10 MHz to 8 GHz
(MA24108A)
10 MHz to 18 GHz
(MA24118A)
10 MHz to 26 GHz
(MA24126A)
Connector Type N(m), 50 Ω Type N(m), 50 Ω Type N(m), 50 Ω Type N(m), 50 Ω
(MA24108/18A)
Type K(m), 50 Ω
(MA24126A)
Dynamic Range -30 to +20 dBm (.001 to
100 mW)
+3 to +51.76 dBm (2 mW to
150 W)
-40 to +23 dBm
(0.1 µW to 200 mW)
-40 to +20 dBm
(0.1 µW to 100 mW)
VBW 100 Hz 100 Hz 100 Hz 50 kHz
Measurand True-RMS True-RMS True-RMS True-RMS,
Slot Power, Burst Average
Power
Measurement Uncertainty ± 0.16 dB
1
± 0.17 dB
2
± 0.16 dB
1
± 0.18 dB
3
Datasheet
(for complete specications)
11410-00414 11410-00483 11410-00424 11410-00504
Notes: 1) Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than -20 dBm with zero mismatch errors.
2) Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load. Measurement
results referenced to the input side of the sensor.
3) Expanded uncertainty with K=2 for power measurements of a CW signal greater than -20 dBm with zero mismatch errors.